Mobile menu

OPT & examination lights

OPT light

GSRIPENTALED 30E

Technical data:

  • Light intensity at 1m distance (Ec): 130.000 Lux 
  • Color temperature: double selection: 4500 / 5000° K 
  • Color rendering index (CRI): 96 Ra  
  • R9: ≥ 90 
  • No. of Leds: 30  
  • Diameter and focus adjustment: Electronic - fixed
  • Light field diameter adjustable: 16 – 28 cm
  • d10 light field diameter where illuminance reaches 10% of Ec: 160 mm  
  • d50 light field diameter where illuminance reaches 50% of Ec: 80 mm 
  • Depth of illumination IEC 60601-2-41 (L1+L2) at 60%: 49 cm 
  • Depth of illumination IEC 60601-2-41 (L1+L2) at 20%: 95 cm 
  • Control of the illuminance (%): 25 - 100 
  • Total radiated energy Ee where the illuminance reaches max level: 299 W/m2
  • Ratio between radiated energy Ee and illuminance Ec: 2,30 mW/m2.lx
  • Radiated UV energy with wavelength less than 400nm: 0,002 W/m2 
  • Average Led life: > 50.000 hours 
  • EndoLED light intensity min-max: 5.000 – 20.000 Lux 
  • Primary alternating voltage (a.c.): 100 - 240 V
  • Secondary continue voltage (d.c): 24 V
  • Frequency: 50 - 60 Hz
  • Absorbed power: 50 VA

 

Mounting options: 

  • Ceiling suspended
  • Wall mounted
  • Mobile on four castors